dc.identifier.citation | Espada Colón, C. A. (2016). Time optimization of gas chromatography-mass spectrometry method for controlled substances analysis. Politechne, 18(1), 58-62. | |
dc.identifier.issn | 1079-7432 | |
dc.identifier.uri | http://hdl.handle.net/20.500.12475/262 | |
dc.description | Design Project Article for the Graduate Programs at Polytechnic University of Puerto Rico | en_US |
dc.description.abstract | An optimization of the
gas chromatography- mass spectrometry
(GC-MS) method was conducted
by establishing a temperature ramp
of 150°C, held by 3 min and raising
at 12°C/min until reaching 300°C,
which from 25.00 min was reduced to
15.50 min the analysis time and identified
qualitatively the Caffeine, Cocaine,
Delta 9 THC, Oxycodone and
Heroin. A validation of the results was
performed to determine the precision,
instrument detection limit (IDL) and
selectivity from six replicates analysis.
For the controlled substances identification
a 200 ppm QC-Standard
Drug mix solution was prepared. A
low variability was observed in the
retention time of each compound.
Instead, the peaks area demonstrates
a high variability from the acceptance
criteria of RSD, especially the Oxycodone
affecting the IDL results. All the
compounds showed a good resolution
with acceptance criteria of R<2, except
Oxycodone. Further measurements
will be necessary to acquire more tight
values to the acceptances criteria. | en_US |
dc.language.iso | en_US | en_US |
dc.publisher | Polytechnic University of Puerto Rico | en_US |
dc.relation.ispartof | Manufacturing Competitiveness Program; | |
dc.relation.ispartofseries | Revista Politechne: Ingeniería; | |
dc.relation.ispartofseries | Design Project Articles Master Degree; | |
dc.relation.haspart | San Juan Campus | en_US |
dc.subject.lcsh | Gas chromatography--Industrial applications | |
dc.subject.lcsh | Mass spectrometry | |
dc.title | Time Optimization of Gas Chromatography-Mass Spectrometry Method for Controlled Substances Analysis | en_US |
dc.type | Article | en_US |
dc.rights.holder | Esta Junta Editorial y la Universidad Politécnica de Puerto Rico hacen constar y reconoce que los autores de los artículos, obras literarias y artísticas publicadas en esta Revista Politechnê, se reservan enteramente los derechos de autor y de publicación de los mismos para los efectos de cualquier ventualidad literaria, publicitaria o de cualquier índole. | en_US |