dc.identifier.citationRosario Rentas, E. (2016). Titanium Grade 36 to Titanium Grade 1 weld on implantable cardioverter defibrillator (ICD) [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.en_US
dc.identifier.urihttp://hdl.handle.net/20.500.12475/588
dc.descriptionDesign Project Article for the Graduate Programs at Polytechnic University of Puerto Ricoen_US
dc.description.abstractFinding cost reduction on medical devices production, a new Feedthrough array design will be implemented in an Implantable Cardioverter Defibrillator, incorporating glassed Feedthrough of Titanium Grade 36 in a ferrule with glass insulator. This design will be used in place of the brazed Feedthrough of Niobium Type 2 in a ferrule with sapphire as insulator utilized in legacy devices. The Titanium Grade 36 Feedthrough will be welded per Resistance Spot Welding to Interconnect Ribbons of Titanium Grade 1. Being the first time that will be used of Titanium Grade 36 as material on Feedthrough, this new weld will be analyzed and characterized to meet the device requirement and obtain the economic impact of the material change with a benefit of $18-20M NPV. The Design for Reliable Manufacturing method demonstrate that a robust characterization design can be development in order to meet the specification as implement the new FT array knowing all effected factors. Key Terms - Feedthrough, Niobium Type 2, Resistance Spot Welding, Titanium Grade 1, Titanium Grade 36.en_US
dc.language.isoen_USen_US
dc.publisherPolytechnic University of Puerto Ricoen_US
dc.relation.ispartofManufacturing Engineering;
dc.relation.ispartofseriesWinter-2016;
dc.relation.haspartSan Juanen_US
dc.subject.lcshMedical instruments and apparatus industry--Management
dc.subject.lcshImplantable cardioverter-defibrillators
dc.subject.lcshCost control
dc.subject.lcshPolytechnic University of Puerto Rico--Graduate students--Research
dc.titleTitanium Grade 36 to Titanium Grade 1 weld on Implantable Cardioverter Defibrillator (ICD)en_US
dc.typeArticleen_US
dc.rights.holderPolytechnic University of Puerto Rico, Graduate Schoolen_US


Ficheros en el ítem

Thumbnail

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem