dc.identifier.citationEspada Colón, C. A. (2014). Time optimization of gas chromatography-mass spectrometry method for controlled substances analysis [Unpublished manuscript]. Graduate School, Polytechnic University of Puerto Rico.en_US
dc.identifier.urihttp://hdl.handle.net/20.500.12475/769
dc.descriptionDesign Project Article for the Graduate Programs at Polytechnic University of Puerto Ricoen_US
dc.description.abstractAn optimization of the gas chromatography- mass spectrometry (GC-MS) method was conducted by establishing a temperature ramp of 150°C, held by 3 min and raising at 12°C/min until reaching 300°C, which from 25.00 min was reduced to 15.50 min the analysis time and identified qualitatively the Caffeine, Cocaine, Delta 9 THC, Oxycodone and Heroin. A validation of the results was performed to determine the precision, instrument detection limit (IDL) and selectivity from six replicates analysis. For the controlled substances identification a 200 ppm QC-Standard Drug mix solution was prepared. A low variability was observed in the retention time of each compound. Instead, the peaks area demonstrates a high variability from the acceptance criteria of RSD, especially the Oxycodone affecting the IDL results. All the compounds showed a good resolution with acceptance criteria of R<2, except Oxycodone. Further measurements will be necessary to acquire more tight values to the acceptances criteria. Key Terms - Controlled substances, GC-MS, Temperature, Validationen_US
dc.language.isoen_USen_US
dc.publisherPolytechnic University of Puerto Ricoen_US
dc.relation.ispartofManufacturing Competitiveness;
dc.relation.ispartofseriesSpring-2014;
dc.relation.haspartSan Juanen_US
dc.subject.lcshGas chromatography--Industrial applications
dc.subject.lcshDrugs of abuse--Analysis
dc.subject.lcshPolytechnic University of Puerto Rico--Graduate students--Research
dc.titleTime Optimization of Gas Chromatography-Mass Spectrometry Method for Controlled Substances Analysisen_US
dc.typeArticleen_US
dc.rights.holderPolytechnic University of Puerto Rico, Graduate Schoolen_US