Finished Goods Failure Reduction in a Wound Closure Medical Device Product Family

Thumbnail
Ver/
Final Article _Ameiris Alvarez (418.3Kb)
Fecha
2013
Autor
Álvarez Vázquez, Ameiris
Metadatos
Mostrar el registro completo del ítem
Resumen
A corrective action and preventive action plan was developed to address the increase of Finished Goods failures in a wound closure medical device product family. Following a DefineMeasure-Analyze-Improve-Control Six Sigma approach, process failure modes were successfully identified and mitigated. Results show Finished Goods nonconformances decreased by 93 percent during the first four months of the action plan effectiveness period increasing the process sigma level from 5.51 to 6.35. Key Terms - Corrective Action, Failure Investigation, Root Cause, Six Sigma.
URI
http://hdl.handle.net/20.500.12475/842
Colecciones
  • Manufacturing Engineering

PRC Repository copyright © 2022  COBIMET, Inc.
Contacto
Theme by 
Atmire NV
 

 

Listar

Todo PRCRComunidades & ColeccionesPor fecha de publicaciónAutoresTítulosMateriasEsta colecciónPor fecha de publicaciónAutoresTítulosMaterias

Mi cuenta

AccederRegistro

Estadísticas

Ver Estadísticas de uso

PRC Repository copyright © 2022  COBIMET, Inc.
Contacto
Theme by 
Atmire NV